Method and apparatus for creating a virtual microscope slide6775402Abstract A method and apparatus are disclosed for constructing a virtual microscope slide comprised of digitally scanned images from a microscope specimen. The digitally scanned images are arranged in a tiled format convenient for viewing without a microscope, and for transferring the tiled images for viewing by another at a remote location. Several original microscope views at a low magnification are digitized and stored as digitized images coherently seamed together to provide an overall virtual, macro image of the specimen at a lower resolution. Several original microscope views at higher magnifications are digitized and stored as digitized images coherently seamed together to provide virtual micro images at higher resolution. A data structure is formed with these virtual macro and micro digitized images along with their mapping coordinates. Preferably, a generic viewing program is also provided in the data structure that allows remote users to manipulate and interpret the tiled images on the user's monitor. Also, the data structure is formed with significantly compressed data so as to be transmitted over low bandwidth channels, such as the Internet, without loss of resolution that would interfere with the analysis at a remotely-located pathologist receiving the data structure over the Internet. The preferred interactive program allows the pathologist to scroll and view neighboring image areas of interest. A marker on the macro image indicates to the user the location of the micro image and assists the user in selecting areas from the macro image to be viewed at higher resolution and magnification. Claims What is claimed is: Description BACKGROUND OF THE INVENTION
TABLE 1
Slidescan.ini
[Header]
x = 278000
y = 142500
1XStepSize = 48143
1YStepSize = 35800
iScannedCount = 37
[Ss1]
x = 181714
y = 142500
[Ss2]
x = 133571
y = 142500
[Ss3]
x = 37285
y = 106700
[Ss4]
x = 85428
y = 106700
[Ss5]
x = 133571
y = 106700
[Ss6]
x = 181714
y = 106700
[Ss7]
x = 229857
y = 106700
[Ss8]
x = 229857
y = 70900
[Ss9]
x = 181714
y = 70900
[Ss10]
x = 133571
y = 70900
[Ss11]
x = 85428
y = 70900
[Ss12]
x = 37285
y = 70900
[Ss13]
x = -10858
y = 70900
[Ss14]
x = -10858
y = 35100
[Ss15]
x = 37285
y = 35100
[Ss16]
x = 85428
y = 35100
[Ss17]
x = 133571
y = 35100
[Ss18]
x = 181714
y = 35100
[Ss19]
x = 229857
y = 35100
[Ss20]
x = 278000
y = -700
[Ss21]
x = 229857
y = -700
[Ss22]
x = 181714
y = -700
[Ss23]
x = 133571
x = -700
[Ss24]
x = 85428
y = -700
[Ss25]
x = 37285
y = -700
[Ss26]
x = -10858
y = -700
[Ss27]
x = -10858
y = -36500
[Ss28]
x = 37285
y = -36500
[Ss29]
x = 85428
y = -36500
[Ss30]
x = 133571
Y = -36500
[Ss31]
x = 181714
y = -36500
[Ss32]
x = 229857
y = -36500
[Ss33]
x = 278000
y = -36500
[Ss34]
x = 278000
y = -72300
[Ss35]
x = 229857
y = -72300
[Ss36]
x = 181714
y = -72300
[Ss37]
x = 133571
y = -72300
Table 2 is a listing of the file FinalScan.ini, which is a listing the X-Y coordinates of the high magnification image tiles scanned and stored.
TABLE 2
FinalScan.ini
[Header]
tPatientID = mda027
tAccession =
tOperatorID = jwb
tTimeOfScan = 8/4/97 1:19:56
PM
1XStageRef = 278000
1YStageRef = 142500
iImageWidth = 752
iImageHeight = 480
1XStepSize = 1590
1YStepSize = 1190
1XOffset = -1900
1YOffset = -400
dMagnification = 40
1AnalysisImageCount = 105
1CalibrationImageCount = 0
[Da0]
x = 214532
y = 65584
[Da1]
x = 212996
y = 65584
[Da2]
x = 211460
y = 65584
[Da3]
x = 209924
y = 65584
[Da4]
x = 208388
y = 65584
[Da5]
x = 206852
y = 65584
[Da6]
x = 205316
y = 65584
[Da7]
x = 203780
y = 65584
[Da8]
x = 214532
y = 64400
[Da9]
x = 212996
y = 64400
[Da10]
x = 211460
y = 64400
[Da11]
x = 209924
y = 64400
[Da12]
x = 208388
y = 64400
[Da13]
x = 206852
y = 64400
[Da14]
x = 205316
y = 64400
[Da15]
x = 203780
y = 64400
[Da16]
x = 214532
y = 63216
[Da17]
x = 212996
y = 63216
[Da18]
x = 211460
y = 63216
[Da19]
x = 209924
y = 63216
[Da20]
x = 208388
y = 63216
[Da21]
x = 206852
y = 63216
[Da22]
x = 205316
y = 63216
[Da23]
x = 203780
y = 63216
[Da24]
x = 214532
y = 62032
[Da25]
x = 212996
y = 62032
[Da26]
x = 211460
y = 62032
[Da27]
x = 209924
y = 62032
[Da28]
x = 208388
y = 62032
[Da29]
x = 206852
y = 62032
[Da30]
x = 205316
y = 62032
[Da31]
x = 203780
y = 62032
[Da32]
x = 214532
y = 60848
[Da33]
x = 212996
y = 60848
[Da34]
x = 211460
y = 60848
[Da35]
x = 209924
y = 60848
[Da36]
x = 208388
y = 60848
[Da37]
x = 206852
y = 60848
[Da38]
x = 205316
y = 60848
[Da39]
x = 203780
y = 60848
[Da40]
x = 214532
y = 59664
[Da41]
x = 212996
y = 59664
[Da42]
x = 211460
y = 59664
[Da43]
x = 209924
y = 59664
[Da44]
x = 208388
y = 59664
[Da45]
x = 206852
y = 59664
[Da46]
x = 205316
y = 59664
[Da47]
x = 203780
y = 59664
[Da48]
x = 214532
y = 58480
[Da49]
x = 212996
y = 58480
[Da50]
x = 211460
y = 58480
[Da51]
x = 209924
y = 58480
[Da52]
x = 208388
y = 58480
[Da53]
x = 206852
y = 58480
[Da54]
x = 205316
y = 58480
[Da55]
x = 203780
y = 58480
[Da56]
x = 180740
y = 82160
[Da57]
x = 179204
y = 82160
[Da58]
x = 177668
y = 82160
[Da59]
x = 176132
y = 82160
[Da60]
x = 174596
y = 82160
[Da61]
x = 173060
y = 82160
[Da62]
x = 171524
y = 82160
[Da63]
x = 180740
y = 80976
[Da64]
x = 179204
y = 80976
[Da65]
x = 177668
y = 80976
[Da66]
x = 176132
y = 80976
[Da67]
x = 174596
y = 80976
[Da68]
x = 173060
y = 80976
[Da69]
x = 171524
y = 80976
[Da70]
x = 180740
y = 79792
[Da71]
x = 179204
y = 79792
[Da72]
x = 177668
y = 79792
[Da73]
x = 176132
y = 79792
[Da74]
x = 174596
y = 79792
[Da75]
x = 173060
y = 79792
[Da76]
x = 171524
y = 79792
[Da77]
x = 180740
y = 78608
[Da78]
Computer 32 can also use the scanned image files to create a self-executing data structure. By compressing the .bmp images to .jpg and adding a dynamic, self-executing program which enables the user to view, reconstruct and manipulate the image tiles, the user can use the data structure as a virtual microscope slide of the original specimen. Preferably, the dynamic, self-executing program is a Java applet, such as shown on FIG. 7B. Computer 32 can provide the slide image data structure 31 directly or via an intranet browser 33 to local viewer 34, or via an Internet server 38. Slide image data structure 37 is shown as being directly accessible from Internet server 38. Alternatively, a user can download the slide image data structure on his own computer 39, use an internet browser 43 and view the reconstructed images. Another alternative is for computer 32 to store the slide image data structure on a CD-rom, Jazz drive or other storage medium. To view slide image data structure 31 or 37, the user, who for example, has acquired the data structure via a CD-rom, first installs the CD-rom in the CD-rom drive of his computer. Then the user opens up a browser or other applications program which can read the Java applet installed on the CD-rom with the image tiles. Note that in some instances no separate browser program may be required. In some case, the CD-rom may include the complete applications program for viewing, reconstructing and manipulating the image tiles. In the instant example, the user will then select the icon or file listing for the slide image data structure and the control program will display the data files. FIG. 2 is a screen view of a system embodying the present invention showing a low magnification image 24 of a specimen on a microscope slide in one window, a high magnification image 26 of a portion of the low magnification image selected by a region marker 30 and a control window 28. FIG. 3 is a view of a display screen of the apparatus embodying the present invention showing the control window 28, a low magnification window 24 having a plurality of high magnification micro image regions 310 delineated therein and a high magnification window 26 including one or more of the micro image regions 310, 314, 316. FIG. 4 is a view of a macro image of an actual breast cancer specimen displayed at 1.25.times. as seen on a computer monitor. FIG. 5 is a view of the grid portion of FIG. 4 outlining a region of interest selected by a pathologist displayed at 40.times. magnification. Recall that region A in FIG. 1A was about 4.8 mm by 3.5 mm. This area creates 752 by 480 pixels of sensed data, or 360,930 pixels of information. Each pixel sends information about its location and the image it sensed to the computer. The computer stores this information in a series of data files (typically .bmp format, but .tif or .gif could also be used). Thus, it can be seen that several more pixels of sensed data are available for viewing on a computer monitor operating at 640 by 480. To view the entire image, the user must scroll through the image tiles. However, scrolling need not be done on a tile, by tile basis. Rather, the user scrolls by pointing to a pixel on the monitor. FIG. 6 is a block diagram showing how the control program locates and scrolls through the stored image tiles. Using the example from FIG. 1a, a complete data structure has been created. When the user loads the data structure (of the microscope slide) into his personal computer or views it from an Internet browser, the control program recreates a bit map of the stored data. The bit map of the entire slide is shown in FIG. 6. Image tile A is also high-lighted. This bit map enables a user to point to or otherwise reference a location on the slide. The X-Y coordinate information specified in the data structure enables X-Y translation of the specific image tiles and specific pixels within the image tile. When the control program first loads the image, because this image file is so large, only a small number of the available tiles are displayed in the active window on the user's monitor. The user uses his mouse or pointing device to scroll through the active window to view the entire macro image. The X-Y coordinate information selected by the mouse translates into specific image tiles or portions therein. The computer takes the mouse pointer information and retrieves the image data from the series of stored tile images and displays them on the monitor for viewing the by user. Because of the large amount of CCD pixel information stored, actual CCD pixel information can be recreated in the viewing window. The entire system operates in a loop, where the user inputs a mouse location, the computer translates the mouse location from the screen coordinates (screen pixels) to the X-Y coordinates on the bit map. Similarly, the user may select the high magnification data images. These are outlined by a dark grid, indicating the areas stored. The user operates the mouse in the same manner as described above. The control program locates the stored X-Y coordinates and retrieves the selected parts of the image, CCD stored pixel by CCD stored pixel. As mentioned above, to save storage space, computer 32 can perform a data compression on each of the image tile files. A preferred data compression is JPEG, which is readily transferred and recognized by most Internet browser programs. Also, JPEG allows flexibility in the amount of data to be compressed, from 20 to 80 percent. FIG. 8 is file listing such as would be seen under Windows 95 file manager showing the data files included in an alternate data structure, one in which the data files have been compressed or converted to JPEG (.jpg) format for a breast cancer specimen. The file index.html (shown in Table 3) is the listing which contains the X-Y coordinate information for these data files. This is the information that is read by the dynamic, self-executing program for viewing, reconstructing and manipulating the image tiles into the macro and micro views.
TABLE 3
index.html
<HTML>
<TITLE>
DCIS_027 - Web Slide
</TITLE>
<BODY>
<APPLET CODE=WebSlide/BliWebSlide.class NAME=DCIS_027
WIDTH=3384 HEIGHT=960 HSPACE=0 VSPACE=0 ALIGN=Middle>
<PARAM NAME = "tPatientID" VALUE = "mda027">
<PARAM NAME = "tAccession" VALUE = " ">
<PARAM NAME = "tOperatorID" VALUE = "jwb">
<PARAM NAME = "tTimeOfScan" VALUE = "8/4/97 1:19:56 PM">
<PARAM NAME = "1XStageRef" VALUE = "278000">
<PARAM NAME = "1YStageRef" VALUE = "142500">
<PARAM NAME = "iImageWidth" VALUE = "752">
<PARAM NAME = "iImageHeight" VALUE = "480">
<PARAM NAME = "1XStepSize" VALUE = "1590">
<PARAM NAME = "1YStepSize" VALUE = "1190">
<PARAM NAME = "1XOffset" VALUE = "-1900">
<PARAM NAME = "1YOffset" VALUE = "-400">
<PARAM NAME = "dMagnification" VALUE = "40">
<PARAM NAME = "iImageCount" VALUE = "105">
<PARAM NAME = "1XSsStepSize" VALUE = "48143">
<PARAM NAME = "1YSsStepSize" VALUE = "35800">
<PARAM NAME = "iScannedCount" VALUE = "37">
<PARAM NAME = "1StartX" VALUE = "278000">
<PARAM NAME = "1StartY" VALUE = "142500">
<PARAM NAME = "Ss1_X" VALUE = "181714">
<PARAM NAME = "Ss1_Y" VALUE = "142500">
<PARAM NAME = "Ss2_X" VALUE = "133571">
<PARAM NAME = "Ss2_Y" VALUE = "142500">
<PARAM NAME = "Ss3_X" VALUE = "37285">
<PARAM NAME = "Ss3_Y" VALUE = "106700">
<PARAM NAME = "Ss4_X" VALUE = "85428">
<PARAM NAME = "Ss4_Y" VALUE = "106700">
<PARAM NAME = "Ss5_X" VALUE = "133571">
<PARAM NAME = "Ss5_Y" VALUE = "106700">
<PARAM NAME = "Ss6_X" VALUE = "181714">
<PARAM NAME = "Ss6_Y" VALUE = "106700">
<PARAM NAME = "Ss7_X" VALUE = "229857">
<PARAM NAME = "Ss7_Y" VALUE = "106700">
<PARAM NAME = "Ss8_X" VALUE = "229857">
<PARAM NAME = "Ss8_Y" VALUE = "70900">
<PARAM NAME = "Ss9_X" VALUE = "181714">
<PARAM NAME = "Ss9_Y" VALUE = "70900">
<PARAM NAME = "Ss10_X" VALUE = "133571">
<PARAM NAME = "Ss10_Y" VALUE = "70900">
<PARAM NAME = "Ss11_X" VALUE = "85428">
<PARAM NAME = "Ss11_Y" VALUE = "70900">
<PARAM NAME = "Ss12_X" VALUE = "37285">
<PARAM NAME = "Ss12_Y" VALUE = "70900">
<PARAM NAME = "Ss13_X" VALUE = "-10858">
<PARAM NAME = "Ss13_Y" VALUE = "70900">
<PARAM NAME = "Ss14_X" VALUE = "-10858">
<PARAM NAME = "Ss14_Y" VALUE = "35100">
<PARAM NAME = "Ss15_X" VALUE = "37285">
<PARAM NAME = "Ss15_Y" VALUE = "35100">
<PARAM NAME = "Ss16_X" VALUE = "85428">
<PARAM NAME = "Ss16_Y" VALUE = "35100">
<PARAM NAME = "Ss17_X" VALUE = "133571">
<PARAM NAME = "Ss17_Y" VALUE = "35100">
<PARAM NAME = "Ss18_X" VALUE = "181714">
<PARAM NAME = "Ss18_Y" VALUE = "35100">
<PARAM NAME = "Ss19_X" VALUE = "229857">
<PARAM NAME = "Ss19_Y" VALUE = "35100">
<PARAM NAME = "Ss20_X" VALUE = "278000">
<PARAM NAME = "Ss20_Y" VALUE = "-700">
<PARAM NAME = "Ss21_X" VALUE = "229857">
<PARAM NAME = "Ss21_Y" VALUE = "-700">
<PARAM NAME = "Ss22_X" VALUE = "181714">
<PARAM NAME = "Ss22_Y" VALUE = "-700">
<PARAM NAME = "Ss23_X" VALUE = "133571">
<PARAM NAME = "Ss23_Y" VALUE = "-700">
<PARAM NAME = "Ss24_X" VALUE = "85428">
<PARAM NAME = "Ss24_Y" VALUE = "-700">
<PARAM NAME = "Ss25_X" VALUE = "37285">
<PARAM NAME = "Ss25_Y" VALUE = "-700">
<PARAM NAME = "Ss26_X" VALUE = "-10858">
<PARAM NAME = "Ss26_Y" VALUE = "-700">
<PARAM NAME = "Ss27_X" VALUE = "-10858">
<PARAM NAME = "Ss27_Y" VALUE = "-36500">
<PARAM NAME = "Ss28_X" VALUE = "37285">
<PARAM NAME = "Ss28_Y" VALUE = "-36500">
<PARAM NAME = "Ss29_X" VALUE = "85428">
<PARAM NAME = "Ss29_Y" VALUE = "-36500">
<PARAM NAME = "Ss30_X" VALUE = "133571">
<PARAM NAME = "Ss30_Y" VALUE = "-36500">
<PARAM NAME = "Ss31_X" VALUE = "181714">
<PARAM NAME = "Ss31_Y" VALUE = "-36500">
<PARAM NAME = "Ss32_X" VALUE = "229857">
<PARAM NAME = "Ss32_Y" VALUE = "-36500">
<PARAM NAME = "Ss33_X" VALUE = "278000">
<PARAM NAME = "Ss33_Y" VALUE = "-36500">
<PARAM NAME = "Ss34_X" VALUE = "278000">
<PARAM NAME = "Ss34_Y" VALUE = "-72300">
<PARAM NAME = "Ss35_X" VALUE = "229857">
<PARAM NAME = "Ss35_Y" VALUE = "-72300">
<PARAM NAME = "Ss36_X" VALUE = "181714">
<PARAM NAME = "Ss36_Y" VALUE = "-72300">
<PARAM NAME = "Ss37_X" VALUE = "133571">
<PARAM NAME = "Ss37_Y" VALUE = "-72300">
<PARAM NAME = "Da0_X" VALUE = "214532">
<PARAM NAME = "Da0_Y" VALUE = "65584">
<PARAM NAME = "Da1_X" VALUE = "212996">
<PARAM NAME = "Da1_Y" VALUE = "65584">
<PARAM NAME = "Da2_X" VALUE = "211460">
<PARAM NAME = "Da2_Y" VALUE = "65584">
<PARAM NAME = "Da3_X" VALUE = "209924">
<PARAM NAME = "Da3_Y" VALUE = "65584">
<PARAM NAME = "Da4_X" VALUE = "208388">
<PARAM NAME = "Da4_Y" VALUE = "65584">
<PARAM NAME = "Da5_X" VALUE = "206852">
<PARAM NAME = "Da5_Y" VALUE = "65584">
<PARAM NAME = "Da6_X" VALUE = "205316">
<PARAM NAME = "Da6_Y" VALUE = "65584">
<PARAM NAME = "Da7_X" VALUE = "203780">
<PARAM NAME = "Da7_Y" VALUE = "65584">
<PARAM NAME = "Da8_X" VALUE = "214532">
<PARAM NAME = "Da8_Y" VALUE = "64400">
<PARAM NAME = "Da9_X" VALUE = "212996">
<PARAM NAME = "Da9_Y" VALUE = "64400">
<PARAM NAME = "Da10_X" VALUE = "211460">
<PARAM NAME = "Da10_Y" VALUE = "64400">
<PARAM NAME = "Da11_X" VALUE = "209924">
<PARAM NAME = "Da11_Y" VALUE = "64400">
<PARAM NAME = "Da12_X" VALUE = "208388">
<PARAM NAME = "Da12_Y" VALUE = "64400">
<PARAM NAME = "Da13_X" VALUE = "206852">
<PARAM NAME = "Da13_Y" VALUE = "64400">
<PARAM NAME = "Da14_X" VALUE = "205316">
<PARAM NAME = "Da14_Y" VALUE = "64400">
<PARAM NAME = "Da15_X" VALUE = "203780">
<PARAM NAME = "Da15_Y" VALUE = "64400">
<PARAM NAME = "Da16_X" VALUE = "214532">
<PARAM NAME = "Da16_Y" VALUE = "63216">
<PARAM NAME = "Da17_X" VALUE = "212996">
<PARAM NAME = "Da17_Y" VALUE = "63216">
<PARAM NAME = "Da18_X" VALUE = "211460">
<PARAM NAME = "Da18_Y" VALUE = "63216">
<PARAM NAME = "Da19_X" VALUE = "209924">
<PARAM NAME = "Da19_Y" VALUE = "63216">
<PARAM NAME = "Da20_X" VALUE = "208388">
<PARAM NAME = "Da20_Y" VALUE = "63216">
<PARAM NAME = "Da21_X" VALUE = "206852">
<PARAM NAME = "Da21_Y" VALUE = "63216">
<PARAM NAME = "Da22_X" VALUE = "205316">
<PARAM NAME = "Da22_Y" VALUE = "63216">
<PARAM NAME = "Da23_X" VALUE = "203780">
<PARAM NAME = "Da23_Y" VALUE = "63216">
<PARAM NAME = "Da24_X" VALUE = "214532">
<PARAM NAME = "Da24_Y" VALUE = "62032">
<PARAM NAME = "Da25_X" VALUE = "212996">
<PARAM NAME = "Da25_Y" VALUE = "62032">
<PARAM NAME = "Da26_X" VALUE = "211460">
<PARAM NAME = "Da26_Y" VALUE = "62032">
<PARAM NAME = "Da27_X" VALUE = "209924">
<PARAM NAME = "Da27_Y" VALUE = "62032">
<PARAM NAME = "Da28_X" VALUE = "208388">
<PARAM NAME = "Da28_Y" VALUE = "62032">
<PARAM NAME = "Da29_X" VALUE = "206852">
<PARAM NAME = "Da29_Y" VALUE = "62032">
<PARAM NAME = "Da30_X" VALUE = "205316">
<PARAM NAME = "Da30_Y" VALUE = "62032">
<PARAM NAME = "Da31_X" VALUE = "203780">
<PARAM NAME = "Da31_Y" VALUE = "62032">
<PARAM NAME = "Da32_X" VALUE = "214532">
<PARAM NAME = "Da32_Y" VALUE = "60848">
<PARAM NAME = "Da33_X" VALUE = "212996">
<PARAM NAME = "Da33_Y" VALUE = "60848">
<PARAM NAME = "Da34_X" VALUE = "211460">
<PARAM NAME = "Da34_Y" VALUE = "60848">
<PARAM NAME = "Da35_X" VALUE = "209924">
<PARAM NAME = "Da35_Y" VALUE = "60848">
<PARAM NAME = "Da36_X" VALUE = "208388">
<PARAM NAME = "Da36_Y" VALUE = "60848">
<PARAM NAME = "Da37_X" VALUE = "206852">
<PARAM NAME = "Da37_Y" VALUE = "60848">
<PARAM NAME = "Da38_X" VALUE = "205316">
<PARAM NAME = "Da38_Y" VALUE = "60848">
<PARAM NAME = "Da39_X" VALUE = "203780">
<PARAM NAME = "Da39_Y" VALUE = "60848">
<PARAM NAME = "Da40_X" VALUE = "214532">
<PARAM NAME = "Da40_Y" VALUE = "59664">
<PARAM NAME = "Da41_X" VALUE = "212996">
<PARAM NAME = "Da41_Y" VALUE = "59664">
<PARAM NAME = "Da42_X" VALUE = "211460">
<PARAM NAME = "Da42_Y" VALUE = "59664">
<PARAM NAME = "Da43_X" VALUE = "209924">
<PARAM NAME = "Da43_Y" VALUE = "59664">
<PARAM NAME = "Da44_X" VALUE = "208388">
<PARAM NAME = "Da44_Y" VALUE = "59664">
<PARAM NAME = "Da45_X" VALUE = "206852">
<PARAM NAME = "Da45_Y" VALUE = "59664">
<PARAM NAME = "Da46_X" VALUE = "205316">
<PARAM NAME = "Da46_Y" VALUE = "59664">
<PARAM NAME = "Da47_X" VALUE = "203780">
<PARAM NAME = "Da47_Y" VALUE = "59664">
<PARAM NAME = "Da48_X" VALUE = "214532">
<PARAM NAME = "Da48_Y" VALUE = "58480">
<PARAM NAME = "Da49_X" VALUE = "212996">
<PARAM NAME = "Da49_Y" VALUE = "58480">
<PARAM NAME = "Da50_X" VALUE = "211460">
<PARAM NAME = "Da50_Y" VALUE = "58480">
<PARAM NAME = "Da51_X" VALUE = "209924">
<PARAM NAME = "Da51_Y" VALUE = "58480">
<PARAM NAME = "Da52_X" VALUE = "208388">
<PARAM NAME = "Da52_Y" VALUE = "58480">
<PARAM NAME = "Da53_X" VALUE = "206852">
<PARAM NAME = "Da53_Y" VALUE = "58480">
<PARAM NAME = "Da54_X" VALUE = "205316">
<PARAM NAME = "Da54_Y" VALUE = "58480">
<PARAM NAME = "Da55_X" VALUE = "203780">
<PARAM NAME = "Da55_Y" VALUE = "58480">
<PARAM NAME = "Da56_X" VALUE = "180740">
<PARAM NAME = "Da56_Y" VALUE = "82160">
<PARAM NAME = "Da57_X" VALUE = "179204">
<PARAM NAME = "Da57_Y" VALUE = "82160">
<PARAM NAME = "Da58_X" VALUE = "177668">
<PARAM NAME = "Da58_Y" VALUE = "82160">
<PARAM NAME = "Da59_X" VALUE = "176132">
<PARAM NAME = "Da59_Y" VALUE = "82160">
<PARAM NAME = "Da60_X" VALUE = "174596">
<PARAM NAME = "Da60_Y" VALUE = "82160">
<PARAM NAME = "Da61_X" VALUE = "173060">
<PARAM NAME = "Da61_Y" VALUE = "82160">
<PARAM NAME = "Da62_X" VALUE = "171524">
<PARAM NAME = "Da62_Y" VALUE = "82160">
<PARAM NAME = "Da63_X" VALUE = "180740">
<PARAM NAME = "Da63_Y" VALUE = "80976">
<PARAM NAME = "Da64_X" VALUE = "179204">
<PARAM NAME = "Da64_Y" VALUE = "80976">
<PARAM NAME = "Da65_X" VALUE = "177668">
<PARAM NAME = "Da65_Y" VALUE = "80976">
<PARAM NAME = "Da66_X" VALUE = "176132">
<PARAM NAME = "Da66_Y" VALUE = "80976">
<PARAM NAME = "Da67_X" VALUE = "174596">
<PARAM NAME = "Da67_Y" VALUE = "80976">
<PARAM NAME = "Da68_X" VALUE = "173060">
<PARAM NAME = "Da68_Y" VALUE = "80976">
<PARAM NAME = "Da69_X" VALUE = "171524">
<PARAM NAME = "Da69_Y" VALUE = "80976">
<PARAM NAME = "Da70_X" VALUE = "180740">
<PARAM NAME = "Da70_Y" VALUE = "79792">
<PARAM NAME = "Da71_X" VALUE = "179204">
<PARAM NAME = "Da71_Y" VALUE = "79792">
<PARAM NAME = "Da72_X" VALUE = "177668">
<PARAM NAME = "Da72_Y" VALUE = "79792">
<PARAM NAME = "Da73_X" VALUE = "176132">
<PARAM NAME = "Da73_Y" VALUE = "79792">
<PARAM NAME = "Da74_X" VALUE = "174596">
<PARAM NAME = "Da74_Y" VALUE = "79792">
<PARAM NAME = "Da75_X" VALUE = "173060">
<PARAM NAME = "Da75_Y" VALUE = "79792">
Referring now to the drawings, and especially to FIGS. 9A, 9B and 10, apparatus for synthesizing low magnification and high magnification microscopic images is shown therein and generally identified by reference numeral 10. The system includes a computer 12 which is a dual Pentium Pro personal computer in combination with a Hitachi HV-C20 video camera 14 associated with a Zeiss Axioplan 2 microscope 16. The computer system 12 is able to receive signals from the camera 14 which captures light from the microscope 16 having a microscope slide 18 positioned on an LUDL encoded motorized stage 20. The encoded motorized stage 20 includes a MAC 2000 stage controller for controlling the stage in response to the computer 12. A microscope slide 18 includes a biological specimen 21 which is to be viewed by the microscope and whose image is to be digitized both at low magnification and at high magnification as selected by a user. The low magnification digitized image is then displayed on a 21 inch Iiyama video display monitor 22 having resolution of 1600 by 1200 to provide display screens of the type shown in FIGS. 1 through 3 including a low magnification image 24, for instance, at 1.25 power, a high magnification image 26, for instance at 40.times. power and a control window or image 28. The low magnification image may have identified therein a region 30 which is reproduced at high magnification in high magnification screen or window 26 so that a pathologist or other operator of the system can review architectural regions of interest in low magnification image 24 and simultaneously view them in high magnification in the high magnification screen or window 26 to determine whether the cells forming a portion of the architectural feature need be examined further for cancer or the like or not. The computer 10 is constructed around a PCI system bus 40 and has a first Pentium Pro microprocessor 42 and a second pentium pro microprocessor 44 connected thereto. The system bus 40 has connected to it a PCI bus 50 and an ISA bus 52. The PCI bus 50 has a SCSI controller 60 connected thereto to send and receive information from a hard disk 62. The hard disk 62 also is coupled in daisy chain SCSI fashion to a high capacity removal disk and to a CD Rom drive 66. The hard disks 62 contains the programs for operating the system for controlling the microscope 16 and for processing the images as well as for doing a quantitative analysis of the selected portions of the histological specimens being viewed on the slide 18. The system bus 40 also has connected to it a random access memory 70 within which portions of the program being executed are stored as well as a read only memory 72 for holding a bootstrap loader as well as portions of the basic input/output operating system. A floppy disk controller 74 is coupled to the system bus 40 and has connected to it a floppy disk drive 76 for reading and writing information to a floppy disk as appropriate. A mouse controller 80 is coupled to the system bus and has a mouse 82 which operates as a pointing device for controlling manipulations on the screen 22 and within the windows 24, 26 and 28. A keyboard controller 90 is connected to the system bus and has a keyboard 92 connected thereto. The keyboard 92 may be used to send and receive alpha numeric signals to other portions of the computer. An audio controller 100 has a plurality of speakers 102 and a microphone 104 connected thereto for audio input and output and is coupled to the system bus 40. A network interface, such as a network interface card 104, is connected to the system bus and can provide signals via a channel 106 to other portions of a network or internet to which the system may be connected. Likewise, signals can be sent out of the system through a modem 110 connected to the ISA bus 52 and may be sent via a channel 112, for instance, to the internet. A printer 116 is connected via a parallel I/O controller 118 to the system bus in order to provide printouts as appropriate of screens and other information as it is generated. A serial I/O controller 122 is connected to the system bus and has connected to it a camera controller 124 which is coupled to CCD sensors 126 in the cameras. The CCD sensors 126 supply pixel or image signals representative of what is found on the slide 18 to an Epix pixci image acquisition controller 130 coupled to the PCI bus 50. The microscope 16 includes a base 140 having a stage 20 positioned thereon as well as an objective turret 142 having a plurality of objectives 144, 146 and 148 thereon. The objective 144, for instance, may be of 1.25.times. objective. The objective 146 may be a 20.times. objective. The objective 148 may be a 40.times. objective. Signals from the camera sensors and controller are supplied over a bus 128 to the image acquisition system where they are digitized and supplied to the PCI bus for storage in RAM or for backing storage on the hard disk 62. When a specimen is on the slide 18 the stage 20 may be manipulated under the control of the computer through a stage controller 160 coupled to the serial I/O controller 122. Likewise, a microscope controller 162 controls aspects of the microscope such as the illumination, the color temperature or spectral output of a lamp 168 and the like. For instance, in normal operation, when a specimen is placed on the slide, specimen slide 18 is placed on the stage 20 in a step 200, as shown in FIG. 14, the processors 42 or 44 send a command through the system bus to cause the serial I/O controller 122 to signal the microscope controller to change magnification to 1.25.times. in a step 202. This is done by rotating the objective turret of the Axioplan 2 microscope to select the objective 144. Likewise, the controller sets the color temperature of the lamp 168, sets a pair of neutral density filter wheels 170 and 172 and sets a field diaphragm 174 for the correct illumination. A condenser diaphragm 176 is also controlled and a color filter wheel 180 may also be controlled to apply the appropriate filter color to the CCD censors 126 in the camera. The entire slide is then scanned in a step 204. The images are tiled and melded together into the overall image 24 supplied on the screen 22 to provide the operator in the step 206 with a visually inspectable macro image of relevant regions of the slide of interest. In order to provide the magnified image, the mouse may be moved to identify a marker segment or region which, for instance, may be a rectangular region which will cause the microscope to change magnification as at step 208 to 4.times., 20.times., 40.times., etc., by rotating the turret to bring the appropriate objective lens system into viewing position. Next the user, in a step 209a, uses the mouse to select the region on the macro image in order to select the micro image to be viewed on the screen 22. In a step 209b a test is made to determine whether the user has commanded continued inspection. If the user has, a test is made in a step 209c to determine if the magnification is to be changed by changing the selected objective. In the event the magnification is to be changed control is transferred to the step 208. If the magnification is to remain unchanged control is transferred to the step 209a. In the event inspection is not to continue the region selected is outlined for higher magnification scan in a step 209d. In a step 209e, a command may be received to scan or acquire the higher magnification image for display in screen 26. The image may then be archived for later analysis, displayed or analyzed immediately. In order to perform the magnification called for in step 208, the overall illumination and control of the microscope will be controlled so that in a step 210 the objective turret 142 will be rotated to place the higher power objective above the slide 18. In a step 212 voltage to the lamp will be changed to adjust the lamp 168 to provide the proper illumination and color temperature as predetermined for the selected objective. In a step 214, the condenser diaphragm 176 will have its opening selected as appropriate to provide the proper illumination for that objective. In a step 216, the filter turret 180 will select the proper light wavelength filter to be supplied to the camera sensors. For instance, a red, blue or green filter, as appropriate, particularly if the specimen has been stained. In a step 218 the field diaphragm 174 will have its opening changed. In a step 220 the neutral density filter wheel 170 will select a neutral density filter and in a step 222 the neutral density filter wheel 172 will also select a neutral density filter. In a step 224 the X, Y and Z offsets will be used for reconstruction of the recorded image at the magnification and in a step 226 the current position will be read from encoders in the stage which are accurate to 0.10 micron. In order to identify the selected region the mouse is moved to that area of the region in a pointing operation in a step 240 as shown in FIG. 14. The mouse may be moved to draw a box around the region selected. In a step 242 the X and Y screen points are computed for the edges of the regions selected and the computed image or pixel points are translated to stage coordinate points in order to control the stage of the microscope. In a step 244 a list of all of the X fields for positioning the stage for the objective is stored in random access memory and may be backed up on the hard disk. The information from the X offsets for the objective and the stage offsets is used as well as the size of the field to position the slide properly under the objective to capture the micro image. When the slide has been positioned properly, as shown in FIG. 15 in a step 250 the stage is positioned for each of the X and Y coordinate values in stage coordinate values and the digitized image is captured by the cameras and stored in RAM and backed up on the hard disk. The image may be then analyzed quantitatively in various manners such as those set forth in the previously-identified United States application. Optionally the image may be stored for archival purposes in a step 254. In order to override the specific control functions that take place as shown in FIG. 12, a screen is provided as shown in FIG. 13 wherein the X-Y step size can be edited, the X, Y and Z offset can be edited, the lamp voltage can be selected, the neutral density filter can be selected as well as the opening of the field diaphragm and several other microscopic characteristics. FIG. 13 is a view of the settings of the microscope objective properties of the Axioplan 2, computer-controlled microscope. The X and Y positioning is specifically carried out as shown in FIG. 16 where the slide 18 is shown with a slide boundary 270, 272, 274 and 276. Stage boundary for limits of the stage travel for purposes of the stage the stage can be moved all the way from an upper left hand corner of travel 276 to a lower right hand corner of travel 280. At the upper left hand bounded corner of travel 278 limits which a signal that the end of travel has been reached and the stage is then translated a short distance 282 in the extra action and a short distance 284 in the Y direction to define the first tile 288 in terms of a reference point 290 at its upper left hand corner. Since the size of the macro image tile 288 is known, the next macro image tile 292 may be placed contiguous with it by moving the stage appropriately and by measuring the location of the stage from the stage in counters without the necessity of performing any image manipulation. The image tiles 288 and 292 may be abutted without any substantial overlap or they may be overlapped slightly, such as a one pixel with overlap, which is negligible insofar as blurring of any adjacent edges of abutted image tiles. The upper left hand corner 300 of the tile 292 defines the rest of 292 and other tiles can be so defined. Micro image tiles can likewise be defined so that they are contiguous but not substantially overlapping, as would interfere with the composite image. This avoids the problems encountered with having to perform extended computations on digital images in a frame storer or multiple frame storage in order to match or bring the images into contiguity without blurriness at the edges of contiguous image tiles. It may be appreciated that the low power image 24 has a plurality of micro images defined therein which are tiled and which are shown in higher magnification as individual tiles 312, 314, 316 and the like. In addition, the region 310 when magnified as shown in the window 26 may exceed the bounds of the window and thus the window may include scroll bars or other means for allowing the image 310 which is larger than the window 26 to be examined from within the window 26. The stage 200 is best seen in FIG. 16A and includes the X and Y stepper motors 279 and 281 with their respective encoders, which provide a closed loop system to give the 0.1 micron accuracy versus the usual 5 or 6 micron accuracy of most microscope stages without a closed loop system. This closed loop system and this very high accuracy allow the abutting of the tile images for both high magnification and low magnification images without the substantial overlap and the time-consuming and expensive software currently used to eliminate the overlap and blurriness at the overlapping edges of adjacent image tiles. With the precisely positioned stage and by using the tiling system described in connection with FIG. 16, where the slide is precisely positioned relative to a center point CP for the slide, and the known position of point 278 is always taken from the same point, the tiles may be positioned precisely in a horizontal row and precisely in vertical rows to reconstruct the macro image and the micro image. This reconstruction is done without the use, as in the prior art, of extensive software manipulation to eliminate overlapping image tiles, horizontally or vertically or the haphazard orientation of image tiles. The present invention also includes the facility for allowing remote observation to occur by being able to couple the system either over a network communication facility to an intranet, for instance via the network interface, or via a modem or other suitable connection, to an internet so that once the image has been scanned and stored in memory on hard disks or other storage, remote users may be able to access the low magnification image as well as the high magnification image and move around within both images to make determinations as to the histological characteristics of the samples. An additional feature of the system includes a plurality of networked workstations coupled to a first computer console 12 having a display screen 22 connected to the microscope 14. Satellite work stations 350 and 352 are substantially identical to the work station 12 including respective computers 354 and 356 coupled to displays 358 and 360. The devices can be manipulated through input devices 360 and 362 which may include a keyboard, mouse and the like. Also a third device can be connected including a work station 370, having a display 372, a computer 374 and an input device 376. Each of the devices is connected over respective network lines 380, 382, 384 to the computer 12 which transmission may be via either net or the like. Each of the different operators at the physically separate viewing stations can locate regions from the view of entire tissue cross sections via a macro view and label the regions for subsequent scanning and/or quantitative analysis. A single operator at the instrument station 12 can locate regions to view the entire tissue cross section. Those regions can be labeled for subsequent scanning and/or quantitative analysis with subsequent review and physically remote viewing stations, for instance, in an operating room or in individual pathologists' signout areas in order to review analysis results while still maintaining and reviewing the entire macro view of the tissue and/or the individual stored images from which the quantitative results were obtained. The viewing stations 350, 352 and 370 can comprise desk top computers, laptops, etc. There is no need for a microscope at the network stations 3150, 352 and 370. In a still further alternative embodiment, remote workstations 400, 402, 404, 406 and 408 may be connected through a server 410 which may be supplied via a packet switched network. The server 410 and may be a hypertext transport protocol based server of the type used for the World Wide Web or may be a telnet type server as used previously in internet remote operation applications. The server 410 communicates via a communications channel 414 with a local computer 416 having a display 418 associated therewith, the local computer 416 being connected to the microscope 420. Each of the remote work stations 400, 402, 404, 406 and 408 may perform the same operations as the stations 350, 352 and 370 although they do it from nearby buildings or even from around the world, thus providing additional flexibility for others to make use of the specimen obtained and being viewed under the microscope 420. In addition, stored images may be disseminated through the server 410 to the remote servers 400 through 408 for further analysis and review. While there has been illustrated and described a particular embodiment of the present invention, it will be appreciated that numerous changes and modifications will occur to those skilled in the art, and it is intended in the appended claims to cover all those changes and modifications which followed in the true spirit and scope of the present invention.
|
Same subclass Same class Consider this |
||||||||||
